Photometry is the measurement of light absorbed in the ultraviolet (UV) to visible (VIS) to infra-red (IR) range. This measurement is used to determine the amount of an analyte in a solution or liquid. Photometers utilize a specific light source and detectors that convert light passed through a sample solution into a proportional electric signal. These detectors may be for example photodiodes, photoresistors, or photomultipliers. Photometry uses Beer–Lambert’s law to calculate coefficients obtained from the transmittance measurement. A correlation between absorbance and analyte concentration is then established by a test-specific calibration function to achieve highly accurate measurements. Photometry is a widely used quantitative analysis in research laboratories to determine the amounts of inorganic and organic compounds in solutions and other liquids. Photometry also has broad industrial applications in the determination of contaminants in drinking and wastewater, analysis of nutrients in soil, food and beverage samples, building material composition, and many other areas. Photometry
The components of a typical photometer include a light source, monochromator, sample, and detector. Light sources can be tungsten-halogen lamps (generally the source of light used for analyses in the visible light range) or LEDs. For measurements in the UV–Visible range a xenon flash lamp may be used. The monochromator filters the light radiated by the light source to allow only a very narrow spectrum to pass. The light then passes through the cuvette or the sample holding cell. Based on the amount of analyte (or a dye derived from it) that is present in the sample solution, some part of the light is absorbed by the solution and the remaining is transmitted. The transmitted light is directed towards the detectors, which produce an electric current proportional to the light intensity. Beer–Lambert law, also known as Beer’s law, states that the quantity of light absorbed by a sample is directly proportional to the concentration of the analyte present and path length of the light through the sample. “Sample” in this context means either the analyte itself (direct measurement) or a dye derived from the analyte (when using reagents or kits). The relationship is described by the formula:Construction and Working of Photometer
Beer–Lambert Law
A = elc where A = absorbance of the sample The spectrophotometer measures the intensity of light before and after passing through a solution and relates it to the transmittance (T) by the following equation: Transmittance(T)=It/Io It and Io are the intensity of light after and before passing through the solution, respectively. The transmittance is related to the absorbance by the equation: Absorbance(A)=−log(T)
l = optical path length (cm)
c = concentration of the analyte (M)
e = molar extinction coefficient (M-1cm-1)
Reflectometry (also known as remission photometry) is a non-destructive analytical technique that uses the reflection of light by surfaces and interfaces to measure characteristics such as color intensity, film thickness and refractive index. As with other photometers, the main elements of reflectometers include a light source, usually long-life LEDs of specific wavelengths that are focused onto a sample surface via a lens system and the reflected light is measured by detectors. Reflectometers are often designed to measure physical characteristics of surfaces like color changes on a test strip. In this approach, a sample can be placed on a test strip and its remission (REM) value can be compared against appropriate controls and standards. As in photometry, the difference in intensity between emitted and reflected light allows a quantitative determination of the concentration of specific analytes. Reflectometry is commonly used in industrial applications as a rapid, sensitive method for quantitating a wide variety of organic and inorganic parameters in water, food, beverages, and environmental samples as well as other diverse uses such as surface analysis of building materials and skin color quantification.Reflectometry
Related Technical Articles
- A transmittance to absorbance table enables fast conversion from transmittance values to absorbance in the lab or in the field.
Transmittance to Absorbance Table
- We are a global leader in the life science industry and have produced test kits to measure numerous analytes. Water Online spoke with us about advances in measuring chemical oxygen demand.
Measuring Chemical Oxygen Demand in Water Treatment Facilities
- Step-by-step accurate reflectometric determination of ammonium in wastewater with Nessler’s reagent or Indophenol blue using Reflectroquant® system and test strips.
Ammonium in Wastewater
- Preparation of a standard solution for COD/chloride
Analytical Method: Analytical Quality Assurance, Standard for COD/Chloride
- Learn about the UV-spectrometric determination of the K268 nm value of olive oil, assessing the UV absorbance at 268 nm, to ensure quality and compliance with regulatory standards.
K268 Olive Oil Commission Regulation (EEC) No 2568/91 Annex IX
- See All (24)
Related Protocols
- Preparation of a standard solution for Hydrogen peroxide
Analytical Method: Analytical Quality Assurance, Standard for Hydrogen peroxide
- Photometric determination with Chromazurole S subsequent to acid mineralisation; DIN decomposition
Analytical Method: Aluminium (total) in effluents
- Photometric determination with Chromazurole S subsequent to acid mineralisation
Analytical Method: Aluminium (total) in effluents, high organic content
- Photometric determination with Chromazurole S subsequent to fusion melting
Analytical Method: Aluminium (total) in lime stone
- Step-by-step method describing the photometric determination of ammonium ions in effluents using Spectroquant® test kits and spectrophotometer.
Ammonium in Effluents
- See All (33)
Find More Articles and Protocols
Related Applications
Drinking Water Testing
Clean drinking water is essential for food and beverage production. This article gives an overview of chemical and microbiological testing of the drinking water for quality and safety.
Water treatment is required to maintain effluent compliance with standards established by regulatory bodies. Learn about the stages of wastewater and process water treatment along with the corresponding analytical testing methods.
Soil and groundwater testing are crucial to avoid adverse impact on the environment and human health. Learn about the importance of solid waste management and methods used for soil and groundwater testing for heavy metals and other environmental pollutants.
Related Product Categories
Explore Spectroquant® solutions: instruments, software, test kits, accessories. From start to finish, ensure rapid, accurate results with user-friendly handling.
Our selection of MQuant® and Reflectoquant® test strips includes a variety of test strips for fast, precise, and low-cost analysis in-the-field or on-site.
Discover and explore our range of MQuant® and Reflectoquant® pH indicator papers, kits, and non-bleeding pH strips for rapid pH measurement easier than ever. Order now.
Browse our colorimetric and titrimetric test kits with brilliant color cards for easy and precise analysis designed for aquaculture, surface water and industrial water testing.
Milli-Q® Benchtop Lab Water Purification Systems
Find the water purification system that fits your lab’s needs. Compact Milli-Q® pure and ultrapure benchtop systems can meet your application, quality, volume, space and budget requirements.
RELATED SERVICES
Spectroquant® Photometer Services
Get support for professional grade maintenance, qualification, and repair to optimize performance and reliability of your Spectroquant® photometer.
Offering high-quality MQuant® and Reflectoquant® test strips, papers as well as wet chemistry items, in private labeling and customized packaging.
Get experienced support for your lab water system and discover digital services that simplify system care. Our experts install, troubleshoot, validate and qualify systems, optimizing performance and lifetime.
Highlights
Webinar: Quickly analyze chemical disinfectant parameters of your production line after disinfection Join our webinar on June 15th to find out how to quickly analyze chemical disinfectant parameters to ensure the safety of your production line after disinfection & the different methods available.
Sign In To Continue
To continue reading please sign in or create an account.
Don't Have An Account?